臨床講演 【MRCP】1 - 関西Gyro Meeting ~For Next Scan

第19回関西Gyro Meeting
市立岸和田市民病院
装置:Ingenia3.0T/R4.1.3
コイル:ds-Torso coil
第19回関西Gyro Meeting
ご施設名:市立岸和田市民病院
B-FFE BH
Tips
T2FS_TRA
B-FFE/BH
Scan time 00:19.5
Geometry
Coil selection
ds-Torso
FOV ×RFOV
350×90%
Matrix × Scan percentage
256×160%
SENSE(P reduction)
Yes(4.0)
slices/thickness/gap
30/5/0.5
orientation/foldover direction Coronal/RL
REST/type
0/Contrast
Scan mode/technique
M2D/FFE
contrast enhancement
balanced
TE/TR
2.1/4.3
Flip angle
80
water-fat shift
Minimum(0.334)
Fat suppression
no
Motion
breath hold
respiratory compensation
Navigator respiratory comp 1
NSA
第19回関西Gyro Meeting
ご施設名:市立岸和田市民病院
AxT2/ssh/BH
Tips
Axに関しては、CTに合わせるため
3.6mm/0.4mm
T2FS_TRA
AxT2/ssh/BH
Scan time 19.6sec
Geometry
Coil selection
FOV (mm)×RFOV
Matrix × Scan percentage
SENSE(P reduction)
slices/thickness/gap (mm)
orientation/foldover direction
REST/type
ds-Torso
350×80%
256×70%
Yes(2.0)
30/3.6/0.4
TRA/AP
0
Contrast
Scan mode/ technique
Fast Imaging mode
TSE factor
startup echo
profile order
DRIVE
TE/TR (ms)
Flip angle
Refocusing control(angle)
MS/SE
TSE
Single-Shot
0
linear
yes
100/shortest
90
No
Motion
respiratory compensation
Flow compensation
NSA
Breath hold
No
1
第19回関西Gyro Meeting
ご施設名:市立岸和田市民病院
2D MRCP
Tips
ステレオ視する為
正面、-15°、+15°を撮像
T2FS_TRA
2D Radial
Scan time 6sec (x3)
Geometry
Coil selection
FOV (mm)×RFOV
Matrix × Scan percentage
SENSE(P reduction)
slices/thickness/gap (mm)
orientation/foldover direction
REST/type
Contrast
Scan mode/ technique
Fast Imaging mode
TSE factor
startup echo
profile order
DRIVE
TE/TR (ms)
Flip angle
Refocusing control(angle)
Motion
respiratory compensation
Flow compensation
NSA
ds-Torso
250×100%
325×70%
No
3/70 radial 15°
COR/RL
0
M2D/SE
TSE
Single-shot
0
linear
no
800/6000
90
No
Breath hold
No
1
第19回関西Gyro Meeting
ご施設名:市立岸和田市民病院
SagT2/ssh/BH
Tips
T2FS_TRA
SagT2/ssh/BH
Scan time 19.3sec
Geometry
Coil selection
FOV (mm)×RFOV
Matrix × Scan percentage
SENSE(P reduction)
slices/thickness/gap (mm)
orientation/foldover direction
REST/type
Contrast
Scan mode/ technique
Fast Imaging mode
TSE factor
startup echo
profile order
DRIVE
TE/TR (ms)
Flip angle
Refocusing control(angle)
Water fat shift
Motion
respiratory compensation
Flow compensation
NSA
ds-Torso
320×80%
256×100%
Yes(2.0)
25/3.0/0.3
Sag/AP
0
M2D/SE
TSE
Single-Shot
0
linear
yes
90/shortest
90
Yes(110)
User defined(1)
Breath hold
yes
1
第19回関西Gyro Meeting
ご施設名:市立岸和田市民病院
CorT2/ssh/BH
Tips
T2FS_TRA
CorT2/ssh/BH
Scan time 19.1sec
Geometry
Coil selection
FOV (mm)×RFOV
Matrix × Scan percentage
SENSE(P reduction)
slices/thickness/gap (mm)
orientation/foldover direction
REST/type
Contrast
Scan mode/ technique
Fast Imaging mode
TSE factor
startup echo
profile order
DRIVE
TE/TR (ms)
Flip angle
Refocusing control(angle)
Water fat shift
Motion
respiratory compensation
Flow compensation
NSA
ds-Torso
320×90%
256×100%
Yes(4.0)
30/3.0/0.3
Cor/RL
0
M2D/SE
TSE
Single-Shot
0
linear
yes
90/shortest
90
Yes(110)
User defined(1)
Breath hold
yes
1
第19回関西Gyro Meeting
ご施設名:市立岸和田市民病院
Ax/Dual/BH
Tips
T2FS_TRA
Ax/Dual/BH
Scan time 18.3sec
Geometry
Coil selection
FOV (mm)×RFOV
Matrix × Scan percentage
SENSE(P reduction)
slices/thickness/gap
orientation/foldover direction
REST/type
Contrast
Scan mode/ technique
Fast Imaging mode
TE/TR (ms)
Flip angle
Water fat shift
Motion
respiratory compensation
Flow compensation
NSA
dS-Torso
320×75%
192×100%
Yes(2.0)
35/3.6/0.4
TRA/AP
2/parallel
MS/FFE
None
Out1.15 in2.3/240
60
Minimum(0.294)
Breath hold
No
1
第19回関西Gyro Meeting
ご施設名:市立岸和田市民病院
Ax/DWI/FB
Tips
T2FS_TRA
Ax/DWI/FB
Scan time 3’09”
Geometry
Coil selection
FOV (mm)×RFOV
Matrix × Scan percentage
SENSE(P reduction)
slices/thickness/gap
orientation/foldover direction
Contrast
Scan mode/technique
Fast Imaging mode
TE/TR
Half scan(factor)
water-fat shift
Fat suppression
DWI mode
b-factor
average high b
Motion
respiratory compensation
NSA
dS-Torso
320×100%
112 ×120%
Yes(2.0)
35/3.6/0.4
Transverse/AP
MS/SE
EPI
70/8750
Yes(0.6)
Minimum
SPAIR inv.100
DWI
0/1000
yes
No
2
第19回関西Gyro Meeting
ご施設名:市立岸和田市民病院
MRCP/3D/Navi
Tips
T2FS_TRA
MRCP/3D/Navi
Scan time 3:39sec
Geometry
Coil selection
FOV (mm)×RFOV
Matrix × Scan percentage
SENSE(P reduction)
slices/thickness/gap (mm)
orientation/foldover direction
REST/type
Contrast
Scan mode/ technique
Fast Imaging mode
TSE factor
startup echo
profile order
DRIVE
TE/TR (ms)
Flip angle
Refocusing control(angle)
Fat Suppression
Motion
Navigator resp.comp
Flow compensation
NSA
dS-Torso
280×100%
320×78.54%
Yes(P:2.0 S:1.0)
90/1.0/0
COR/RL
0
3D/SE
TSE
Single-shot
20
rev_linear/Y
yes
550/shortest
90
Yes(120)
SPIR
trigger and track
yes
1